Correcting magnification error in foveal avascular zone area measurements of optical coherence tomography angiography images with estimated axial length
Journal article   Open access   Peer reviewed

Correcting magnification error in foveal avascular zone area measurements of optical coherence tomography angiography images with estimated axial length

D.D.C.S. Dutt, S. Yazar, J. Charng, D.A. Mackey, F.K. Chen and D.M. Sampson
Eye and Vision, Vol.9(1), Art. 29
2022
pdf
axial length.pdfDownloadView
Published (Version of Record) Open Access
url
Free to Read *No subscription requiredView

Abstract

Background To generate and validate a method to estimate axial length estimated (ALest) from spherical equivalent (SE) and corneal curvature [keratometry (K)], and to determine if this ALest can replace actual axial length (ALact) for correcting transverse magnification error in optical coherence tomography angiography (OCTA) images using the Littmann-Bennett formula. Methods Data from 1301 participants of the Raine Study Gen2-20 year follow-up were divided into two datasets to generate (n = 650) and validate (n = 651) a relationship between AL, SE, and K. The developed formula was then applied to a separate dataset of 46 participants with AL, SE, and K measurements and OCTA images to estimate and compare the performance of ALest against ALact in correcting transverse magnification error in OCTA images when measuring the foveal avascular zone area (FAZA). Results The formula for ALest yielded the equation: ALest = 2.102K − 0.4125SE + 7.268, R2 = 0.794. There was good agreement between ALest and ALact for both study cohorts. The mean difference [standard deviation (SD)] between FAZA corrected with ALest and ALact was 0.002 (0.015) mm2 with the 95% limits of agreement (LoA) of − 0.027 to 0.031 mm2. In comparison, mean difference (SD) between FAZA uncorrected and corrected with ALact was − 0.005 (0.030) mm2, with 95% LoA of − 0.064 to 0.054 mm2. Conclusions ALact is more accurate than ALest and hence should be used preferentially in magnification error correction in the clinical setting. FAZA corrected with ALest is comparable to FAZA corrected with ALact, while FAZA measurements using images corrected with ALest have a greater accuracy than measurements on uncorrected images. Hence, in the absence of ALact, clinicians should use ALest to correct for magnification error as this provides for more accurate measurements of fundus parameters than uncorrected images.

Details

UN Sustainable Development Goals (SDGs)

This output has contributed to the advancement of the following goals:

#3 Good Health and Well-Being

Source: InCites

Metrics

35 File views/ downloads
75 Record Views

InCites Highlights

These are selected metrics from InCites Benchmarking & Analytics tool, related to this output

Collaboration types
Domestic collaboration
International collaboration
Citation topics
1 Clinical & Life Sciences
1.36 Ophthalmology
1.36.813 Vision Disorders
Web Of Science research areas
Ophthalmology
ESI research areas
Clinical Medicine
Logo image