Logo image
Development and characterisation of a high-sensitivity X-ray CT polymer gel dosimeter
Journal article   Open access   Peer reviewed

Development and characterisation of a high-sensitivity X-ray CT polymer gel dosimeter

Jonathon Lexor Lumley, Pejman Rowshanfarzad, Mounir Ibrahim, Mario Djukelic and David J Henry
Physical and engineering sciences in medicine
2025
PMID: 40632460
pdf
X-ray CT3.20 MBDownloadView
CC BY V4.0 Open Access

Abstract

Sensitivity X-ray CT Polymer gel dosimetry Optimisation PASSAG-N
Polymer gel dosimeters have shown potential for clinical 3D dosimetry; however, their use has been limited due to low sensitivity and reliance on scarcely available magnetic resonance imaging. This study aimed to optimise a PASSAG (Poly AMPS Sodium Salt And Gelatin) polymer gel dosimeter for X-ray computed tomography, to enhance its clinical feasibility. The total monomer concentration was increased to improve sensitivity, and different cosolvents were tested to enhance the limited solubility of N, N'-methylenebisacrylamide, the crosslinker. n-propanol was identified as the optimal cosolvent, allowing for an 18.3% monomer concentration, 30% crosslinker to total comonomer mass gel, at a 3:7 cosolvent-to-water ratio. The optimised formulation, PASSAG-N (PASSAG- n-propanol), consisted of 54.4% w/w deionised water, 23.3% n-propanol, 12.8% 2-acrylamido-2-methylpropane sulfonic acid sodium salt, 5.5% N, N'-methylenebisacrylamide, 4.0% gelatin, and 0.089% (4.65mM) tetrakis (hydroxymethyl)phosphonium chloride. The dosimeter was irradiated within a standard timeframe to assess its sensitivity, and theoretical calculations confirmed its equivalence to water, soft tissue, brain, and muscle. Compared to a cosolvent-free formulation, PASSAG-N exhibited a 250% increase in Hounsfield unit (HU) change, demonstrating enhanced sensitivity. The optimised gel showed a linear response over a 1-12 Gy dose range, with an average sensitivity of 1.072 ± 0.041 HU Gy⁻¹ and a dose resolution ≤ 0.31 Gy, making it a promising alternative for clinical X-ray computed tomography-based dosimetry. This study highlights the potential of PASSAG-N as a highly sensitive and potentially practical polymer gel dosimeter for clinical applications.

Details

UN Sustainable Development Goals (SDGs)

This output has contributed to the advancement of the following goals:

#3 Good Health and Well-Being

Source: InCites

Metrics

36 File views/ downloads
8 Record Views

InCites Highlights

These are selected metrics from InCites Benchmarking & Analytics tool, related to this output

Collaboration types
Domestic collaboration
Citation topics
1 Clinical & Life Sciences
1.147 Prostate Cancer
1.147.289 Radiotherapy Innovations
Web Of Science research areas
Engineering, Biomedical
Radiology, Nuclear Medicine & Medical Imaging
ESI research areas
Engineering
Logo image