Logo image
Improved apparatus and procedures for isopiestic studies at elevated temperatures
Journal article   Open access   Peer reviewed

Improved apparatus and procedures for isopiestic studies at elevated temperatures

G. Hefter, P.M. May, S.L. Marshall, J. Cornish and I. Kron
Review of Scientific Instruments, Vol.68(6), pp.2558-2567
1997
pdf
Improved_apparatus_and_procedures_for_isopiestic.pdfDownloadView
Published (Version of Record) Open Access
url
Free to Read *No subscription requiredView

Abstract

A significantly improved apparatus for the isopiestic measurement of thermodynamic activities at elevated temperatures is described. Details of the new design and of an analysis of the propagation of the experimental errors are presented. The apparatus was tested by measurements of calcium chloride against sodium chloride at 353.15 K. Accuracies of better than 0.1% in the osmotic coefficient are routinely achieved. Data are also given for the osmotic coefficient of aqueous sodium hydroxide solutions at 353.15 K as a function of molality (up to 13 mol/kg), measured with respect to both sodium chloride and calcium chloride solutions as reference electrolytes.